Spectrophotometer Models 7310 / 7315 Models 7310 and 7315 use icon driven software and have an improved navigation system for easy and intuitive usability. These instrument have measurement modes for absorbance, % transmittance, concentration, scanning, kinetics and quantitation. The 7310 and 7315 are easy to use and are ideal for routine testing in clinical, veterinary, pharmaceutical and QC laboratories. -Small footprint -Autologging capabilities -Press to read xenon lamp (7315) -Result and method storage to USB memory stick -PC software included -3 year warranty
Supplied with: Mains lead, pack 100 disposable cuvettes, 10 x 10mm cell holder, 2GB USB memory stick, interface cable, PC Application Software on CD-ROM and operating instructions.
Accessories available include an automatic 8 cell changer, Peltier, sipper pump, combined sipper Peltier pump, adjustable path length cuvette holders and micro-cuvette holders. Cuvettes and lamps are available on request. Specifications Optical system: Single beam Wavelength Range:
Resolution: Accuracy: Bandwidth: 320 to 1000nm (7310) 198 to 1000nm (7315) 1nm ± 2nm 5nm Transmittance Range: Resolution: Stray light: Accuracy: 0 to 199.9% 0.1% <0.5%T ± 1% Absorbance: -0.3 to 2.500 Concentration: Range: Resolution: Calibration: -300 to 9999 Selectable 1, 0.1, 0.01 or 0.001 Blank with a single standard or factor Quantitation: Range: Resolution: Calibration: Curve fit algorithms: -300 to 9999 Selectable 1, 0.1, 0.01 or 0.001 Blank with up to 6 standards Linear regression, interpolation, quadratic, quadratic through zero or linear regression through zero Kinetics display: Graphical rate of change, initial and final absorbance/%T and calculated concentration value against standard and factor Spectrum:
Scan Data Internal: Analysis: 320 to 1000nm (7310) 198 to 1000nm (7315) 1, 2 or 5nm selectable Absorbance and wavelength of peaks and troughs Outputs: USB, analogue and serial RS232, optional integrated printer Supply requirements: 230V/50Hz
Type 7315 |